Calibration and Adjustment

Calibration and adjustment

  |   Spectroscopy

Calibration

The calibration coefficients for the device, which are specific to each device, are included with the device and are entered into the software during installation and setup. To import these numbers into the software, after installing the software, import the numbers from this path and click the Save option. Then click the Stop/Start option on the main page twice for these changes to take effect.

 

Set up >> Software >> Calibration

 

Device setup

In order to use the spectrophotometer, it must be calibrated and set to its optimal state before any analysis is perform

 

For spectrum stability, the deuterium-tungsten lamp should be on for approximately 20 minutes before starting thespectrum acquisition

 

To get the appropriate scope, you need to enter the correct integration time. By increasing the integration time (default 2-10) in both the visible and ultraviolet regions, the intensity increases. These values should be chosen so as not to saturate the detector

 

By adjusting the Smoothing values (default 3) and Average values (default 3-10), the spectrum can be smoothed and stabilized, and the signal-to-noise ratio can be increased